看了顯微鏡溫度控制器的用戶又看了
虛擬號將在 180 秒后失效
使用微信掃碼撥號
顯微鏡溫度控制器, 顯微細(xì)胞培養(yǎng)控制器 H401-T-CONTROLLER
配合顯微鏡使用,讓觀察中的細(xì)胞環(huán)境進(jìn)行實時調(diào)控;
選擇合適的配置和型號,可控制細(xì)胞觀察環(huán)境的溫度、濕度、光線、氣體濃度等等多種參數(shù)。
2-Channel Temperature Controller with 4.3'' touch screen interface compatible with all H401 heating devices. One controller can control independently up to two items.
External temperature sensor (included) and self calibration routines allows easy calibration of the heating devices for the highest accuracy on sample temperature.
On board data logging and download procedures to USB drives allow to record temperature as a function of time and to log calibration events and alarms.
The controller is equipped with password protected routines to re-calibrate the sensors against certified thermometers.
Features
T - range: from ambient to 60°C
T - accuracy on specimen: ± 0.3°C
External T - sensor (included).
On-board Data logging and download routines to USB drive.
On-board logging of calibration and alarm events.
Alarm buzzer and External Alarm Connector
One controllers can control independently up to two devices.
Self calibration routines allow to improve temperature accuracy.
Password protected procedure for periodic calibration against certified thermometer
Integrated in third party software: Nikon NIS Elements; Metamorph; Micromanager
H401-T-PAD
The Heated Flat Metal Plate can be positioned on any surface (Table Top, Cabinet work surface, etc) to create a temperature controlled flat surface.
Features
T range: 27 to 45°C
T sensor accuracy: ±0.1°C
T accuracy on specimen: ±0.3°C, after self calibration, if room temperature remains within ±1°C
Plate calibration can be performed in the lab by the operator
Dimensions:
174x234 mm H401-T-PAD-SMALL
297x210 mm (A4) H401-T-PAD-MEDIUM
420x300 mm (A3) H401-T-PAD-LARGE
H401-GLASS TABLE
The ITO glass table realizes a flat and transparent surface at controlled temperature. Ideal for stereo microscopes.
It requires H401-T-CONTROLLER.
Features
T range: ambient to 60°C
T sensor accuracy: ±0.1°C
T accuracy on specimen: ±0.3°C, after self calibration, if room temperature remains within ±1°C
4.3'' touch screen interface
On-board data logging and download routines to USB drive
Plate calibration can be performed in the lab by the operator
Glass thickness: 2 mm
Adjustable height:
from 31 to 37 mm
from 75 to 105 mm
請關(guān)注玉研儀器的更多相關(guān)產(chǎn)品。
如對產(chǎn)品細(xì)節(jié)和價格感興趣,敬請來電咨詢!
暫無數(shù)據(jù)!