此款手腕式振蕩器已改進(jìn)用于滿足現(xiàn)行美國藥典(USP)中二甲基硅油消泡性能測試。
技術(shù)參數(shù):
半徑距離:13.3±0.4cm
容器數(shù)量:2 x 2
振蕩頻率:300Osc/min
幅度:10o
**負(fù)載:10 lbs
外型尺寸:20 x 19 x 18
電源:220V 50Hz
重量:68 lbs
選配:
附:二甲基硅油(Simethicone)消泡性能美國藥典測試方法 Defoaming Activity Foaming solution: 10 mg/mL of octoxynol 9 in water Sample solution: 4 mg/mL of Simethicone in tertiary butyl alcohol. Shake vigorously.[Note—Warm slightly, if necessary, to dissolve. ] Analysis[Note—For each test, use a clean, unused, 250-mL glass jar. ] Add, dropwise, 500 µL of the Sample solution to a clean, unused, cylindrical 250-mL glass jar, fitted with a 50-mm cap, containing 100 mL of the Foaming solution. Cap the jar, and clamp it in an upright position on a wrist-action shaker. Using a radius of 13.3±0.4 cm (measured from the center of the shaft to the center of the bottle), shake for 10 s through an arc of 10 at a frequency of 300 Osc/min; 30 strokes/min. Record the time required for the foam to collapse. The time, in s, for foam collapse is determined at the instant the first portion of foam-free liquid surface appears, measured from the end of the shaking period. Acceptance criteria: The defoaming activity time does not exceed 15 s.